Chemistries CLAROS                 800-227-4224

Solitax t-line sc Turbidity Immersion Probe with Wiper, including a SC200 Controller, PVC

Solitax t-line sc Turbidity Immersion Probe with Wiper, including a SC200 Controller, PVC
Product #: 2983400
USD Price:
$5,390.00
Qty in Cart:0

Compatible for Hach sc controller platform

The provided Hach SC200 controller displays and processes clearly the measurement data.

Certified Turbidity measurement according to DIN EN ISO 7027

Turbidity measurement performance certified to MCERTS regulation and according to DIN EN ISO 7027.

Automatic wiper cleaning system

Elimination of surface films or particles that can affect performance quality by a proven automatic wiper cleaning system.

No calibration required

Long-term calibration stability since time-consuming multipoint calibrations with dilution series are replaced by a single correction factor.

Any two Solitax sc sensors can be installed using one Hach SC200 Controller

What's in the box?

Turbidity probe, controller, user manual

Specifications

Accuracy: Turbidity up to 1000 NTU;
without calibration < 5% of the measured value ±0.01 NTU;
with calibration < 1% of the measured value ±0.01 NTU;
Automatic Wiper: Yes
Cable Length: 10 m (optional extension cables available)
Calibration Method: Turbidity: Formazin or STABLCAL Standard (at 800 NTU); Suspended Solids: Sample specific, based on gravimetric TSS analysis with a correction factor procedure;
Certifications: CE, MCERTS
Controller: SC200
Diameter Sensor: 60 mm
Flow: Max. 3 m/s (the presence of air bubbles affects the measurement)
Includes: Turbidity probe, controller, user manual
Length: 200 mm
Maintenance Interval: 1 h/month
Material: Optics Carrier and Sleeve: PVC black; Wiper Arm: stainless steel 1.4581; Wiper Rubber: silicone rubber (standard) Optional: Viton (LZX578); Wiper Shaft: stainless steel 1.4571;
Max. cable length Sensor-to-Controller: 100 m
Measuring Principle: Infrared Duo scattered light technique for color-independent turbidity measurement, Turbidity in accordance with DIN EN ISO 7027 / TS equivalent DIN 38414
Model: Solitax t-line sc
Mounting: Immersion
Operating Temperature Range: 0 - 40 °C
Parameter: Turbidity
Pressure Range: PVC: < 1 bar or < 10 m
Range: Turbidity: 0.001 - 4000 NTU
Repeatability: Turbidity: < 1 %
Response Time T90: 1 - 300 s adjustable
Weight: 0.52 kg