Chemistries CLAROS                 800-227-4224

Solitax highline sc Turbidity and Suspended Solids Insertion Probe with Wiper, including a SC200 Controller, Stainless Steel

Solitax highline sc Turbidity and Suspended Solids Insertion Probe with Wiper, including a SC200 Controller, Stainless Steel
Product #: 2983900
USD Price:
$10,314.00
Ships within 2 weeks
Qty in Cart:0

Compatible for Hach sc controller platform

The provided Hach sc200 controller displays and processes clearly the measurement data.

Certified measurements according to DIN EN ISO 7027 and DIN 38414

Turbidity measurement performance certified to MCERTS regulation and according to DIN EN ISO 7027. The Suspended solids measurement is in accordance with DIN 38414.

Automatic wiper cleaning system

Elimination of surface films or particles that can affect performance quality by a proven automatic wiper cleaning system.

No calibration required

Long-term calibration stability since time-consuming multipoint calibrations with dilution series are replaced by a single correction factor

For in pipe use only!

An appropriate mounting assembly for insertion application is required.

What's in the box?

Turbidity & Suspended Solids probe, controller, adapter, flange and insertion mounting, user manual

Specifications

Accuracy: Turbidity: Turbidity up to 1000 NTU:
without calibration < 5% of the measured value ±0.01 NTU
with calibration < 1% of the measured value ±0.01 NTU
Automatic Wiper: Yes
Cable Length: 10 m (optional extension cables available)
Calibration Method: Turbidity: Formazin or STABLCAL Standard (at 800 NTU).
Suspended Solids: Sample specific, based on gravimetric TSS analysis with a correction factor procedure.
Certifications: CE certified to EN 61326-1, EN 61326/A1, EN 61326/A2, EN 61010-1, MCERTS
Controller: SC200
Detection Limit: Turbidity
Suspended Solids
Diameter Sensor: (diameter x length)
Insertion sensor: 60 mm x(2.4 x)
Flow: Max. 3 m/s (the presence of air bubbles affects the measurement)
Includes: Turbidity & Suspended Solids probe, controller, adapter, flange and insertion mounting, user manual
Length: 315 mm
Lower Limit of Detection (LOD): Turbidity
Suspended Solids
Maintenance Interval: 1 h/month
Material: Optics Carrier and Sleeve: stainless steel 1.4571; Wiper Arm: stainless steel 1.4581; Wiper Rubber: silicone rubber (standard) Optional: Viton (LZX578); Wiper Shaft: stainless steel 1.4571
Max. cable length Sensor-to-Controller: 100 m
Measuring Principle: Infrared Duo scattered light technique for color-independent
turbidity measurement Turbidity in accordance with
DIN EN 27027 / TS equivalent DIN 38414
Model: Solitax highline sc
Mounting: Fixed to the rim of the tank
  Insertion
Operating Temperature Range: 0 - 40 °C
Parameter: Suspended Solids, Turbidity
Pressure Range: Stainless steel insertion sensor: <6 bar or <60 m
Range: TSS-content: 0.001 - 500 g/L; 0.001 - 500,000 mg/L (the upper range is dependent on the sample property)
Turbidity: 0.001 - 4000 NTU
Repeatability: TSS content: < 3 %
Turbidity: < 1 %
Response Time T90: 1 - 300 s adjustable
Units: Units of Measure
Turbidity
Suspended Solids: User selectable—g/L, mg/L, ppm, or % solids
Weight: 1.38 kg